• Product Search
  • Certificates of Analysis

Home > LABORATORY INSTRUMENTS > Laboratory Instrument >

Pesticide Residue Rapid Tester

CATEGORY :

lab instrument

USAGE :

For organophosphorus and carbamate detection.

SPECIFICATION :

8/16 channel

CODE :

hknc

  • DETAIL
  • MSDS
Application:
 
NC Series Pesticide Residue Tester is a rapid, accurate, convenient pesticide residue test device. It is developed according to national newest standard method for food test and inspection and rapid test technique is introduced in its design. This device can be wildly used in organophosphorus and carbamate detection in vegetables and fruits, such as agricultural base, orchard, agricultural market, vegetable distribution center, supermarkets and so on. Besides it also applicable in agricultural products inspection officials, government bureaus, health quarantine who has authority in food safety inspection.
 
Features:
 
1. Good performance and high sensitivity.
2. Ergonomic design
4.3 inch LCD display, streamline exterior with mirage black, elegant and durable.
3. Super bright LED light source, low energy consumption, fast response. 
4. Imported photoelectric transducer, it has high sensitivity, overmatch national standard requirements. 
5. Using loop automatic light calibration system, avoid light intensity fluctuation caused by environment, ensuring light source in stable optimal state.
6. Thermostatic system, avoid divergence, ensuring long time steadiness.
7. Reagent has heat stability, no problem in conservation and transportation during hot weather.
8. Less consumption of sample and reagent, no second pollution.
9. Has 8 or 16 testing units, every unit equipped with individual light source, sample container and detector.
10. Automatic calibrate to 100%, not need manual calibration normally.
11. With self-checking function, automatically display problem on screen, helping operator do maintenance and repair.
12. Large volume, can store up to 780000 items data.
13. Freely query and delete data and print.
14. With USB interface, can connect to computer for computer control of the device and conduct testing, data processing and analysis etc.
15. Inside high storage battery can work up to 24 hours.
16. Optional ODM Inverter for mobile inspection van.
 
Specifications:

 Items  Parameters
 Channel number  8/16
 Inhibitory rate   0~100%
 Absorbance   0~1.6
 Absorbance drift   ≤0.1%/3min
 Absorbance accuracy  ± 1%
 Linear error  ± 0.5%
 Repeatability error  ± 0.5%
 Inhibitory rate  error  ± 2%
 Time per test  1~9 min
 Screen  Full color 4.3inch (Resolution 480×272)
 Size  300×200×130mm
 Battery  6400mA, work up to 24 hours

Pesticide and detection limit

 Pesticide  Detection limit (mg/kg)
 DDVP  0.1
 Parathion  1.0
 Phoxim  0.3
 Methamidohpos  2.0
 Malathion  4.0
 Dimethoate  3.0
 Omethoate  0.8
 Isofenphos-methyl  5.0
 Methomyl  0.1
 Carbosulfan   0.05
 Dipterex  0.2
 Carbofuran  0.05